Mindray ESR-2010 Erythrocyte Sedimentation Rate Analyzer
Technical Specifications:
- Area of application: Blood Sedimentation Rate Analysis
- ESR tubes: 9 * 120 mm, 1.28ml Vacuum / Non – vacuum
- ESR Loading Capacity: 10 samples maximum
- Loading Pattern: Random Access (V4.0) ; 10 samples batch (V6.0)
- Duration: 20 samples per hour
- Analysis time: 30 minutes (V4.0) and 30 & 60 minutes (V6.0)
- Results: in Westergren mm / h (by interpolation)
- Test tube: 1.1ml ~ 1.28 ml blood (non – vacuum ESR tube)
- Temperature correction: autocorrected results to 18 ° C (Manley Table)
- Principle of measurement: infrarot (IR) detection, transmission reading initially and during 30/30 & 60 minutes of sedimentation Reading resolution: + /- 0,2 mm ;
- Result resolution: + /- 1 mm Measurement range: 1 – 140 mm / h *
- Reproducibility: C.V < 5% Display: LCD with backlight interface: LPT
- Standard operating temperature of the printer: 15 °C to 30 °C
- Dimensions (w x h x d): 22 cm x 21.6 cm x 17.5 cm Weight: 2.5 kg
- External power supply: 110 – 220V AC
- ESR 2010 V6.0 (Batch Analyzer) provides 1 hour and 2 hours of ESR results in 30 minutes and 60 minutes simultaneously,
- automatically adjusted to 18 ° C.